Revision as of 17:56, 30 January 2021

Regulation, Specification, or Guidance Requirement
ASTM E1578-18 E-6-1 13.1 The system shall be able to trigger an instrument event after a definable number of uses of that instrument.
 


 
ASTM E1578-18 E-6-2 13.2 The system shall be able to automatically take an instrument offline when an instrument's calibration or maintenance date passes without calibration or maintenance taking place.
 


 
ASTM E1578-18 E-6-3 13.3 The system should allow auto samplers and other robotic systems to be controlled via an application programming interface or web services.
 


 
ASTM E1578-18 E-6-4 13.4 The system should accurately record instrument information, personnel names, dates, and times relating to entered results or determinations.
 


 
ASTM E1578-18 E-6-5 13.5 The system should be able to accept the results uploaded from an interfaced instrument.
 


 
ASTM E1578-18 E-6-6 13.6 The system should be capable of sending sample sequences (e.g., control samples, standards) to an instrument that is bidirectionally interfaced.
 


 
ASTM E1578-18 E-6-7 13.7 The system should provide a means to interface with simple laboratory instruments using technologies such as RS-232, TCP/IP, and USB.
 


 
ASTM E1578-18 E-6-8 13.8 The system should support industry communication standards like Specification E1394 and Health Level 7 for bidirectional communication with laboratory instrumentation software.